SENSOR
Technology for Human life
SENSOR
Technology for Human life
Galvanic cell (lead-oxygen battery)
0~25% Oxygen
Applications
• Ideal for portable instrument
• Industry safety (ambient air monitoring appllications)
• Combustion efficiency analysis
• Etc.
Performance Characteristics |
|
---|---|
Sensor Type | Galvanic cell (lead-oxygen battery) |
Measurement Range |
0~25% Oxygen |
Maximem Overload | 30% Oxygen |
Output Signal |
90±20㎂ |
T90 Response Time | <15 sec for 90% response |
Zero current (offset) |
<0.6 % vol O2 |
Linearity |
Linear |
Recommended Load Resistor |
100 ohms |
Environmental | |
Temperature Range Continuous |
-20℃ to +50℃ |
Pressure Range |
800 to 1200 mbar |
Operating Humidity Range |
5% to 95% Rh |
Life Time |
|
Long Term Output Drift | ≤5% per annum |
Recommended Storage Temp |
0℃to 20℃ |
Expected Operating Life |
≥24 months in air |
Standard Warranty |
24 months from the date of dispatch |
Intrinsic Safety Data | |
Maximum current in normal operation (pure O2) | 0.01A |
Maximum o/c Voltage (10 to 100% O2) | 0.9V |
Maximum s/c Current (10 to 100% O2) | 0.5A |